2.7.2 Installing the Headspace Sampling System

Required Parts

The following items are needed to install the HSS:

Headspace instrument

Y-cable power splitter

Transfer line

Carrier Gas

Charged Battery

Procedure

1

Install a fully charged battery into the HAPSITE.

2

Unplug the AC to DC power converter power supply from the HAPSITE. See

Figure 2-21

.

Figure 2-21 Power Supply to HAPSITE from AC To DC Power Converter

3

Connect the single end of the Y-cable (see

Figure 2-22

) to the power supply

connector removed in step 2.

4

Plug one end of the Y-cable into the HAPSITE 24 V(dc) connector.

5

Plug the other end of the Y-cable into the back of the HSS. See

Figure 2-22

.