3 adjusting the ion energy, Section 7.6.3, adjusting the ion, Section 7.6.3 – INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 273

Advertising
background image

7 - 31

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

7.6.3 Adjusting the Ion Energy

1

Prior to performing this procedure, follow the steps outlined in

section 7.6.1,

Adjusting Base Peak Gain, on page 7-26

and

section 7.6.2, Adjusting

Resolution, on page 7-29

.

2

Check the Actual Percentage column:

2a

If the Actual Percentage is above the acceptable range, (red box appears to
the right of the column, indicating the mass response is too high) the ion energy
needs to be reduced. Lower the ion energy by increments of 5 at a time.

2b

If the Actual Percentage is below the acceptable range, (red box appears to
the left of the column, indicating the mass response is too low) the ion energy
needs to be increased. Raise the ion energy by increments of 5 at a time.

Figure 7-35 Ion Energy Needs to be Increased

NOTE: When adjusting ion energy, there are low and high limits for each mass.

Make sure manual adjustments stay within the limits shown in

Table 7-1

.

Table 7-1 Manual Adjustments Guidelines

MASS

IE low

IE high

50 90 170

55 90

170

69

90 170

Advertising