7 - 28

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

5a

If yes, then the BPG does not need adjustment. Proceed to

Step 8

.

5b

If no, the BPG will need to be adjusted. Proceed to

Step 6

.

6

To raise the Base Peak Gain, increase the value of the EM Voltage by 25
volts
. To lower the BPG, decrease the value of the EM Voltage by 25 volts.

7

Check the Base Peak Gain value to see if the BPG is now in the range (0.8 to
2.0)
.

7a

If the Base Peak Gain is still out of range, repeat

Step 6

until the value comes

into range.

7b

If the Base Peak Gain is in range, continue with

Step 8

.

8

Check the Tune and Mass Calibration Status panel to see if any of the masses
are highlighted in red. A red box indicates the mass is High or Low. See

Figure

7-32

.

Figure 7-32 BPG Good with Mass 50 Low

8a

If all mass percentages are within limits, then the instrument is tuned. Proceed
to

section 7.6.2, Adjusting Resolution, on page 7-29

.

8b

If any mass percentages are High or Low (red boxes), then proceed to
Adjusting Resolution.