7 - 21

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

Figure 7-22 Tune and Mass Calibration Status Menu

Mass . . . . . . . . . . . . . . . . . . . . . . . . The mass number of the peak. When a mass

is selected to display, the mass immediately
below and above are also displayed.

Target Resolution. . . . . . . . . . . . . . Target resolution at 10% peak height.

Actual Resolution. . . . . . . . . . . . . . Measured resolution at 10% peak height.

Resolution. . . . . . . . . . . . . . . . . . . . Resolution value; can be used to input a

change in resolution value.

Ion Energy. . . . . . . . . . . . . . . . . . . . Ion energy value; can be used to input a

change in Ion energy value.

Position (DAC Value) . . . . . . . . . . . Current DAC setting for mass position.

Scan Width . . . . . . . . . . . . . . . . . . . Displays the points measured per AMU.

Min OK Target Percentage . . . . . . Displays the minimum target percentage

required for the mass peak to meet the OK
LOW criteria.

Min Target Percentage. . . . . . . . . . Displays the minimum target percentage

required for the mass peak to meet OK
criteria. Will turn red if actual percentage is
below the minimum percentage.

Actual Percentage . . . . . . . . . . . . . Displays the actual measured target

percentage.

Max Target Percentage . . . . . . . . . Displays the maximum target percentage

required for the Mass Peak to meet OK
criteria. Will turn red if actual percentage is
above the maximum percentage.