INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 226

Advertising
background image

5 - 20

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

Figure 5-25 Headspace Needle

6

When powered on or taken out of Extended Standby (refer to

section 3.1 on

page 3-1

or

section 3.8.1 on page 3-63

), the HAPSITE will automatically start

preparing the start up Headspace method. If the method that the HAPSITE
begins preparing is not the desired one, refer to

Section 5.3, Selecting a New

Method, on page 5-7

.

6a

If running a Headspace method that requires a Concentrator, the HAPSITE will
run a Concentrator Cleanout. Its progress will be shown by a bar graph. If the
Concentrator Cleanout is not successful, refer to

Section 3.1.3, Concentrator

Cleanout Failure, on page 3-11

.

7

As part of the preparation, the HAPSITE will heat all necessary components,
check pressures and run an AutoTune if necessary. Progress of the preparation
will be shown as a bar graph on the Laptop screen. If AutoTune fails, refer to

Section 7.4, AutoTune Failure, on page 7-6

. When the final results message is

displayed, click OK. See

Figure 5-26

.

Figure 5-26 AutoTune Complete

8

Click the RUN METHOD icon. Click on the GCMS folder. Click on the folder of
the desired configuration and then click on the method folder. See Steps 2-4 of

section 5.3 on page 5-7

.

9

Gently press the Headspace needle into the 40 mL vial that contains the
blank/sample. See

Figure 5-25

.