5.6.1 Quick Reference SOP — GC/MS Mode with HSS

1

Double-click the Plus IQ software icon.

2

Verify that the desired sample configuration (i.e., Sample Loop or
Concentrator) is installed.

3

Power on or take the HAPSITE out of Extended Standby.

4

Depending upon the method, the HAPSITE may initiate a Concentrator
cleanout. If needed, see

Section 3.1.2, Concentrator Options (CONC

OPTIONS), on page 3-8

. See

Section 5.3, Selecting a New Method, on page

5-7

if the start up method is not the desired method.

5

Following a 20 minute sample temperature equilibration in the HSS oven,
gently press the Headspace needle into the 40 mL sample vial.

6

When the system has finished preparing, click the RUN METHOD icon and
select the desired method. Refer to Steps 2-4 of

section 5.3 on page 5-7

.

7

Click the Run button in the pop-up window.

8

When the screen prompts Collecting Sample Now and Sampling Is Done, no
action is required from the user.

9

When the run is complete, a METHOD FINISHED prompt will appear.

10

When prompted, place the Headspace needle in the purge vial and press OK.

11

To analyze the data, see

Chapter 9, Data Review

.