HAPSITE Smart Plus Operating Manual
11.4.2 SituProbe
WARNING
Only trained users should create methods. Choosing the
incorrect sampling device and/or temperatures may
result in inaccurate data.
To create a SituProbe method, choose the SituProbe Sample Input Device. This
will bring up another temperature setting window for setting the SituProbe Oven
and Transfer Line temperatures. See
Figure 11-11 SituProbe Parameters
SituProbe Temperatures (C)
Oven . . . . . . . . . . . . . . . . . . . . . . . . The target temperature the Oven must
achieve before starting data acquisition. This
setting is not available when the SituProbe is
disabled.
Transfer Line. . . . . . . . . . . . . . . . . . The target temperature the Transfer Line
must achieve before starting data
acquisition. This setting is not available when
the SituProbe is disabled.
Probe . . . . . . . . . . . . . . . . . . . . . . . . SituProbe temperature settings before
starting the method. This setting is not
available when the SituProbe is disabled.