HAPSITE Smart Plus Operating Manual
10
Replace the chain clamp insuring that the nut points to the left. Use the
torque wrench and tighten to 70 inch /pounds or 8 Nm. See
Figure 16-33
.
Figure 16-33 Torquing Chain Clamp
CAUTION
Do not connect the black plug from the new NEG Pump
to the black socket at this time. This will avoid any
accidental heating of the NEG Pump.
11
Place the old NEG Pump into the protective container with the o-ring
assembly and place back in the box. See
Figure 16-34
.
Figure 16-34 Old NEG Pump