HAPSITE Smart Plus Operating Manual

14

To confirm the Survey results with a GC/MS run, ANALYZE can be touched or
ANALYZE RUN can be pushed during the Survey run.

NOTE: It is advised to begin an Analyze run either while a peak is being

displayed or Survey has been run for the full two minutes without
detecting a compound.

15

If an Analyze method is not going to be run, touch STOP to stop the sampling
process and automatically save the data. See

Figure 3-41

.

Figure 3-41 ANALYZE and STOP

16

A METHOD FINISHED message will appear when the Survey method has
ended. See

Figure 3-42

.

NOTE: The total time required for a Survey analysis is typically less than 3

minutes.

Figure 3-42 Survey Method Finished