INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 117
Advertising
3 - 23
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
14
To confirm the Survey results with a GC/MS run, ANALYZE can be touched or
ANALYZE RUN can be pushed during the Survey run.
NOTE: It is advised to begin an Analyze run either while a peak is being
displayed or Survey has been run for the full two minutes without
detecting a compound.
15
If an Analyze method is not going to be run, touch STOP to stop the sampling
process and automatically save the data. See
Figure 3-41
.
Figure 3-41 ANALYZE and STOP
16
A METHOD FINISHED message will appear when the Survey method has
ended. See
Figure 3-42
.
NOTE: The total time required for a Survey analysis is typically less than 3
minutes.
Figure 3-42 Survey Method Finished