Metrohm viva 1.1 (ProLab) User Manual

Page 134

Advertising
background image

3.6 Run subwindow

■■■■■■■■■■■■■■■■■■■■■■

122

■■■■■■■■

viva 1.1 (for process analysis)

Sample type

Selection of sample type.

Selection

Sample | Standard | Intercept

Default value

Sample

Sample position

Position of the sample on the rack. This number is used to move to the
sample position with the command MOVE. For this, the Sample posi-
tion
option must be set for the parameter Move under Target.

Input range

1 - 999 (Increment: 1)

Default value

1

Sample amount

Numerical value for the sample amount for which the standard result is
calculated (see Glossary, page 1057). Only numbers, decimal point, +, –
and e may be entered in this field. The limit value check does not take
place until the start of the determination.

NOTE

If a value is entered or imported into this field, then the time of the
data entry and the data source (manual, 'Name of the barcode
reader'
or 'Name of the import file') are additionally stored auto-
matically, placed in the determination and identified in the information
subwindow as entry date and data source.

NOTE

If a negative value is entered into this field, then this will also be calcu-
lated as a negative in the formula editor.

Entry

Number with 10 digits at most

Default value

1.0

Sample size unit

Unit of the sample amount. Any text can be entered into this field.

Entry

16 characters

Selection

L | mL | µL | g | mg | µg | pieces

Default value

mL

Advertising