5 add sample dt, 1 add sample dt - overview, Add sample dt – Metrohm viva 1.1 (ProLab) User Manual

Page 470

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

458

■■■■■■■■

viva 1.1 (for process analysis)

Dosing rate

Rate (volume per time unit) of expulsion. The maximum rate depends on
the cylinder volume of the dosing unit used and on the diameter of the
tubing at the selected dosing port.

Input range

0.01 - 166.00 mL/min

Default value

2.00 mL/min

Selection

maximum

CAUTION

The maximum permitted dosing rate of 4 mL/min may not be exceeded
for capillaries with a diameter < 1 mm.

Filling rate

Rate (volume per time unit) of filling. The maximum rate depends on the
cylinder volume of the dosing unit used and on the diameter of the tubing
at the selected fill port.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

5.5.2.4.5

ADD SAMPLE DT

5.5.2.4.5.1

ADD SAMPLE DT - Overview

Dialog window: Method

ADD SAMPLE DT Properties... ADD SAMPLE DT -

'Command name'

Command for the addition of small sample volumes for the calibra-
tion method DT (Dilution Titration) for the determination of suppressor in
electroplating baths.

Appearance
The command has the following appearance:

Parameters
The parameters for the command ADD SAMPLE DT are set in the fol-
lowing dialog window:

ADD SAMPLE DT - Properties

Advertising