3 result view - results overview – Metrohm viva 1.1 (ProLab) User Manual

Page 330

Advertising
background image

4.5 Determination overview

■■■■■■■■■■■■■■■■■■■■■■

318

■■■■■■■■

viva 1.1 (for process analysis)

4.5.2.5.5.3

Result view - Results overview

Tab: Database

Determination overview Determinations Repro-

cess...

Reprocess Result view Results overview

All automatically generated results for each VA measuring command and
each defined substance as well as all user-defined results are listed on the
Results overview tab. Details on the automatically generated results are
displayed on the Results tab (see Chapter 4.8.3, page 367) and details
on the user-defined results on the User-defined results (see Chapter
4.8.4, page 368)
tab.

Results for MLAT
For Calibration method = MLAT and Sample type = Sample, the fol-
lowing results will be displayed for each VA measuring command and
each substance:

'Command name','Substance name'.Concentration

Concentration of the substance in the sample.

'Command name','Substance name'.Concentration.ASD

Absolute standard deviation of the concentration of the substance in the
sample.

'Command name','Substance name'.Concentration.RSD

Relative standard deviation of the concentration of the substance in the
sample.

No results will be shown for Calibration method = MLAT and Sample
type
= Intercept.

Results for DT
For Calibration method = DT and Sample type = Sample, the follow-
ing results will be displayed for each VA measuring command and each
substance:

'Command name','Substance name'.Concentration

Concentration of the substance in the sample.

'Command name','Substance name'.Concentration.ASD

Absolute standard deviation of the concentration of the substance in the
sample.

'Command name','Substance name'.Concentration.RSD

Relative standard deviation of the concentration of the substance in the
sample.

Advertising