Metrohm viva 1.1 (ProLab) User Manual

Page 639

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1 (for process analysis)

■■■■■■■■

627

Command variables
The following command variables are generated by the command CPVS
in the method run and can be used in formulas under the designation
'Command name.Variable identification':

Identification

Description

BSY

Command status:

1 = BUSY, HOLD or ERROR

0 = READY

invalid (variable not available) = Command has
never been started

DBL

Total duration for the processing of the command in
s

ELT

Variable for electrode test:

0 = Electrode test has not been performed

1 = Electrode test has been performed once

2 = Electrode test has been performed twice

3 = Electrode test has been performed three times

FIN

Command status:

1 = Command has ended at least once

0 = Command has not yet ended

invalid (variable not available) = Command has
never been started

HR

Highest current measuring range used:

1 = 224 mA | 3 = 20 mA | 5 = 2 mA | 7 = 200 µA |
9 = 20 µA | 11 = 2 µA | 13 = 200 nA | 15 = 20 nA |
17 = 2 nA | 19= 200 pA

LR

Lowest current measuring range used:

19= 200 pA | 17 = 2 nA | 15 = 20 nA | 13 = 200 nA
| 11 = 2 µA | 9 = 20 µA | 7 = 200 µA | 5 = 2 mA |
3 = 20 mA | 1 = 224 mA

NMP

Number of measuring points in measuring point list

RPM

Stirring rate in min

–1

STY

Type of stop with which the command was stopped:
1 = finished normally, 0 = stopped manually
with
[Quit] or [Stop]