5.6 Evaluation subwindow

■■■■■■■■■■■■■■■■■■■■■■

666

■■■■■■■■

viva 1.1 (for process analysis)

Jump to a new, empty standard, at which time the column number is
increased by +1. Altered data is adopted.

Option of entering the desired column number to which the program
jumps directly when [Enter] is pressed.

Buttons

[Close]

With this button, the data modified in the dialog window for the dis-
played standard solution is not adopted and the dialog window is closed.

[OK]

With this button, the data modified in the dialog window for the dis-
played standard solution is adopted and the dialog window is closed.

5.6.5

Evaluation - Calibration

5.6.5.1

Calibration - Overview

Subwindow section: Method

Evaluation - Calibration

Application
The way that the sample content for the substances is determined from
the peak data is defined in the Evaluation - Calibration section of the
evaluation. Only one calibration method can be selected for each method;
this calibration method is then valid for all available VA measuring com-
mands.

Tabs

General

Calibration curves

5.6.5.2

Calibration - General

Tab: Method

Evaluation - Calibration General

This tab contains the basic settings for the calibration.

Calibration method

Selection of the method with which calibration is to take place.

Selection

no calibration | MLAT | DT | RC | LAT

Default value

no calibration