2 meas opt - general/hardware, General/hardware – Metrohm viva 1.1 (ProLab) User Manual

Page 532

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

520

■■■■■■■■

viva 1.1 (for process analysis)

5.5.2.7.4.2

MEAS Opt - General/Hardware

Tab: Method

MEAS Opt Properties... MEAS Opt - 'Command name' Gen-

eral/Hardware

Command name

Name of the command.

Entry

25 characters

The general parameters for the instrument are defined on this tab.

Device

Device name

Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.

Selection

'Device name' | not defined

Default value

not defined

not defined
In this case, the device name must be assigned by the user at the start
of the method.

Device type

Display or selection of the instrument type. If an instrument is selected in
the Device name field, the Device type field can no longer be edited
and the instrument type belonging to the instrument is displayed.

If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.

Selection

Avantes Spectrometer | 089 Photometer

Default value

Avantes Spectrometer

Sensor
This section is displayed only when Device type = 089 Photometer.

Temperature measurement

Type of temperature measurement.

Selection

continuous | off

Default value

continuous

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.

Advertising