2 set total volume - properties, 5 calibration commands, 1 calibration command - overview – Metrohm viva 1.1 (ProLab) User Manual

Page 488

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

476

■■■■■■■■

viva 1.1 (for process analysis)

Command variables
The following command variables are generated by the command SET
TOTAL VOLUME
in the method run and can be used in formulas under
the designation 'Command name.Variable identification':

Identification

Description

BSY

Command status:

1 = BUSY, HOLD or ERROR

0 = READY

invalid (variable not available) = Command has
never been started

FIN

Command status:

1 = Command has ended at least once

0 = Command has not yet ended

invalid (variable not available) = Command has
never been started

5.5.2.4.10.2

SET TOTAL VOLUME - Properties

Dialog window: Method

SET TOTAL VOLUME Properties... SET TOTAL VOL-

UME - 'Command name'

Command name

Name of the command.

Entry

25 characters

Total volume

Volume

Quantity of the current total volume in the measuring vessel, e.g. after
automatic rinsing or after a manual exchange of solution.

Input range

0.000 - 1,000.000 mL

Default value

0.000 mL

5.5.2.5

Calibration commands

5.5.2.5.1

Calibration command - Overview

Menu item: Method

Insert New command... Calibration

The following calibration commands can be selected:

Advertising