4 Database

viva 1.1 (for process analysis)

■■■■■■■■

369

Variables

on | off (Default value: off)

Shows the variables used in the formula.

NOTE

Variables that have not been created in the run are not displayed and
lead to invalid results.

Calculation of the concentration of a substance:

NOTE

The calculation does not apply to Dilution Titration.

Concentration(sample) = concentration(measuring cell) * (cell volume /
sample amount) * (dilution volume / analysis volume)

Concentration(sample)

Concentration of the analyte in
the sample

Variable = CONC

Concentration(measuring cell)

Concentration of the analyte in
the measuring cell at the moment
of sample measurement

Variable = CONCM

Cell volume

Sum of the volumes of all solu-
tions added to the measuring cell
until the measurement of the
sample.

Variable = VAR{1}.VTOT

Sample amount

Volume of the sample in mL

Dilution volume

Volume to which the sample
amount is filled up with ultrapure
water

Analysis volume

'x' mL of the dilution volume,
which is pipetted into the measur-
ing cell