Metrohm viva 1.1 (ProLab) User Manual

Page 580

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

568

■■■■■■■■

viva 1.1 (for process analysis)

Device

Device name

Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.

Selection

'Device name' | not defined

Default value

not defined

not defined
In this case, the device name must be assigned by the user at the start
of the method.

Device type

Shows the instrument type.

Selection

IO Controller

Default value

IO Controller

Control parameters

Controller description

Selection of the controller output which is to be controlled.

Selection

'Controller name' | not defined

Default value

not defined

Controller type

Shows the type of the selected controller output.

Selection

Direct controller | Two-point controller | PID
controller
| PWM controller

Setpoint value

Setpoint value to which the control is to be set.

This parameter is not displayed when Controller type = Direct control-
ler
.

Input range

-214783647.000 - 214783647.000 'Mapping
unit of the analog input or of the calculated
value'

Default value

'empty' or default value from the selected
controller 'Mapping unit of the analog input
or of the calculated value'

Control variable

Shows the designation of the analog input port or of the calculated value
(formula entry).