5 Method

viva 1.1 (for process analysis)

■■■■■■■■

449

Parameters
The parameters for the ADD AUX command are configured in the follow-
ing dialog window:

ADD AUX - Properties

Command variables
The following command variables are generated by the ADD AUX com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':

Identification

Description

BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

CYL

Cylinder volume of the dosing unit used for the com-
mand

FIN

Command status:

1 = Command has ended at least once

0 = Command has not yet ended

not valid (variable not available) = Command
has not yet been started

VOL

Currently dosed volume in L (Number)

5.5.2.4.2.2

ADD AUX - Properties

Dialog window: Method

ADD AUX Properties... ADD AUX - 'Command

name'

Command name

Name of the command.

Entry

25 characters

Auxiliary solution

Solution

Entry of a solution name, selection of an auxiliary solution from the solu-
tion table or selection of a standard solution from the method or from the
solution table. The selection of a solution from the solution table is neces-
sary only if the solution is to be added automatically with a dosing device