5 Method

viva 1.1 (for process analysis)

■■■■■■■■

403

The series end track is executed in the run only once at the end of the last
determination of a series after the main track. This track can be used, e.g.,
to set sample changers to the required state at the end of a series.

NOTE

No voltammetry commands can be inserted in a series end track (see
Chapter 5.5.2.11.1, page 606)
.

5.5.1.2.6

Exit track

Subwindow: Method

An exit track is a track that can be created manually by the user with
Insert

New track... Exit track or with the icon. It starts with the

EXIT command (see Chapter 5.5.2.9.7, page 589) and ends with the
END command. Only one exit track can be inserted in a given method.

The exit track is executed in the run when all normal tracks and VA
tracks
as well as the main track have been completed or when the
determination is manually canceled.