5 properties - digital outputs, Properties - digital outputs – Metrohm viva 1.1 (ProLab) User Manual

Page 959

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Devices

viva 1.1 (for process analysis)

■■■■■■■■

947

Action
Only active and editable for the Type = Level detector.

If the fill level is too low and the level detector is therefore active, one of
the following actions will automatically be carried out at the start test:

Selection

Record message | Display message | Cancel
determination

Default value

Display message

Record message
The message that the fill level detector on the device is active will auto-
matically be saved in the determination.
Display message
A message is displayed and you can select whether you want to con-
tinue the determination or cancel it. The determination is continued
with [Yes], with [No] it is canceled.
Cancel determination
The ongoing determination will be automatically canceled. The follow-
ing message must be confirmed with [OK].

7.11.5

Properties - Digital outputs

All recognized outputs are automatically listed on the Digital outputs
tab. The parameters for each digital output and the default values to be
used in the Methods program part can be set.

Port

Shows the internal designation of the port.

Port description

Shows the port designation.

Port specification

Shows the port specification.

Type

Shows the port type.

Default value

Shows the default value which the output takes directly after starting and
exiting the program. The default value is also set after an error or when
canceling the method.

Parameter 1

The meaning of Parameter 1 depends on the Type set.