4 lift, 1 lift - overview, Lift – Metrohm viva 1.1 (ProLab) User Manual
Page 445
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
433
Swing angle
Angle by which the robotic arm is to be swung relative to the current
position. This parameter is visible only when Swing = Relative angle.
Input range
-180.0 - 180.0 °
Default value
10.0 °
Parameters
Swing rate
Selection of the speed at which the robotic arm rotates.
Input range
10 - 55 °/s
Default value
55 °/s
5.5.2.3.4
LIFT
5.5.2.3.4.1
LIFT - Overview
Dialog window: Method
▶ LIFT ▶ Properties... ▶ LIFT - 'Command name'
Command for moving to predefined positions at a sample changer.
Instruments
This command can be executed with the following instruments:
Sample changer: 814, 815, 858, 919
Appearance
The command has the following appearance:
Parameters
The parameters for the LIFT command are configured in the following
dialog window:
■
Command variables
The following command variables are generated by the LIFT command in
the method run and can be used in formulas under the designation 'Com-
mand name.Variable identification':