3.9 Curves subwindow

■■■■■■■■■■■■■■■■■■■■■■

172

■■■■■■■■

viva 1.1 (for process analysis)

ASD

Absolute standard deviation of the substance concentration in the sample.

RSD

Relative standard deviation of the substance concentration in the sample.

Regression data

Function

Function that was applied for the calculation of the calibration curve.

Coefficient of determination calculated from the calibration function.

Evaluation quantity

Evaluation quantity (area or height) used for the evaluation.

Curve type

Curve type used for the calculation of the calibration function.

Weighting

Indication whether the weighting 1/evaluation quantity² was used.

Calibration data

Calibration factor

Concentration of the standard in the measuring vessel for the evaluation
ratio (only for calibration method DT).

Intercept value

The intercept value is the charge that is obtained from the measurement
of the intercept solution (see Glossary, page 1053) (only for calibration
method LAT).

Electrolyte value

The electrolyte value is the charge obtained from the measurement of the
electrolyte solution (see Glossary, page 1053) (only for calibration method
RC).

Date

Date on which the calibration was carried out.

Determination ID

Determination ID of calibration.