3 cvs - pretreatment, Pretreatment – Metrohm viva 1.1 (ProLab) User Manual

Page 629

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1 (for process analysis)

■■■■■■■■

617

Stirrer

Stirring rate

Number of revolutions per minute of the stirrer.

Input range

0 - 3000 min

⁻¹ (Increment: 100)

Default value

2500 min

⁻¹

Hydrodynamic measurement

on | off (Default value: on)

If this check box is activated, then stirring will continue during the sweep.

Only editable for Sensor type = RDE, otherwise disabled.

5.5.2.11.2.3

CVS - Pretreatment

Tab: Method

CVS Properties... Pretreatment

Command name

Name of the command.

Entry

25 characters

The parameters for the pretreatment can be set on this tab.

Stirring time

Waiting time during which the solution is stirred in the measuring cell.

Input range

0.0 - 9999.9 s (Increment: 0.1)

Default value

5.0 s

Potential 1

Potential that is applied to the electrodes during the Waiting time 1.
This potential can be used e.g. as cleaning potential for the electrochemi-
cal cleaning of solid-state electrode surfaces that have been contaminated
with reaction products of electrode redox processes.

Input range

–5.00 - 5.00 V (Increment: 0.001)

Selection

off

Default value

off

Waiting time 1

Waiting time during which the Potential 1 potential is applied to the
electrodes.

Only editable for Potential 1

off, otherwise inactive.

Input range

0.0 - 9999.9 s (Increment: 0.1)

Default value

0.0 s