Metrohm viva 1.1 (ProLab) User Manual

Page 685

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1 (for process analysis)

■■■■■■■■

673

Linear interpolation

y = CALC 0 + CALC 1

x

≥ 2

At least three points are required for being able to generate a nonlinear or
quadratic calibration. A suitable curve fit (quadratic or nonlinear) must be
placed on the points so that the calibration curve is available in the curve
display and/or the calibration curve can be calculated. Fig. 1 shows the
first three points of the calibration, which lie upon a line. This curve can-
not be evaluated with a nonlinear or quadratic curve type. For the purpo-
ses of evaluation, the calibration curve must intersect the line with the
evaluation ratio and may not lie parallel to it.

Figure 1

Fig.: 1

An additional, fourth point is required for the calibration so that the non-
linear or quadratic function can be placed on the points and extrapolated
to the intersection point with the evaluation ratio (see Figure 2, page
673)
.

Figure 2

Fig.: 2

NOTE

The curve type for the standard solution and the curve type for the
sample need not be identical.

Advertising