5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

594

■■■■■■■■

viva 1.1 (for process analysis)

ID1 - ID16

on | off (Default value: off)

If a check box is activated, then the respective sample identifications ID1 -
ID16
will be requested. If other names are assigned for these variables in
the START command, then these names will be displayed in the Sample
data request
dialog window as field titles.

NOTE

Text is selected as the default type for the sample identifications ID1 -
ID16
. If you wish to enter numbers for these method variables that are
to be used for later calculations, then these variables must be switched
to the Number type on the Sample data variables tab in the START
command (see Chapter 5.5.2.9.2.4, page 580).

Sample type

on | off (Default value: off)

If this check box is activated, then the Sample type will be requested.

Sample position

on | off (Default value: off)

If this check box is activated, then the position of the sample on the rack
will be requested.

Sample amount

on | off (Default value: off)

If this check box is activated, then the Sample amount will be requested
(see Glossary, page 1057).

Sample amount unit

on | off (Default value: off)

If this check box is activated, then the Sample amount unit will be reques-
ted.

Analytical volume

on | off (Default value: off)

If this check box is activated, then the Analytical volume will be requested
(see Glossary, page 1046).