Metrohm viva 1.1 (ProLab) User Manual

Page 167

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Workplace

viva 1.1 (for process analysis)

■■■■■■■■

155

Confirmation necessary to accept data during determination

on | off (Default value: off)

If this check box is activated, then data is imported during a determination
only if the Sample data request dialog of a REQUEST command is
opened or if a line from the sample table is open for editing in the Edit
line - Sample table 'Name'
dialog window.

Import from file
on | off (Default value: off)

If this check box is activated, then external data from a file (e.g. via LIMS)
will be imported into a sample table. When this is the case, a check is
made to determine whether the specified import file is available when the
sample table is opened, when the sample table is loaded into the working
sample table and also periodically (every 10 s) when the sample table is
open (although not when the dialog window Edit line is open). If the
import file is available, then the data from this file is automatically inserted
at the end of the sample table. The import file is deleted after each
import.

NOTE

No matter the number of data fields defined on the Display tab, the
import file must always contain for each line the entirety of sample data
in the format Method;ID1;…;ID16;Sample type;Sample posi-
tion;Sample amount;Unit;Analysis volume;Dilution volume
.

Import file

File that is to be imported. The file can be selected in the Select files to

import dialog window with

.

Entry

File name

Import format

Selection of the import format for the import of data from a file.

Selection

*.csv | *.csv (Unicode)

Default value

*.csv

*.csv
Format for the import of CSV files with Western European ASCII char-
acter sets.
*.csv (Unicode)
Format for the import of CSV files with Unicode character sets.

Advertising