Metrohm viva 1.1 (ProLab) User Manual

Page 576

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

564

■■■■■■■■

viva 1.1 (for process analysis)

Mode of operation
At the time of the command, the statuses of all of the analog inputs are
entered and saved in command variables.

Parameters
The parameters for the ANALOG IN command are set on the following
tabs:

General
Parameters for instrument.

Stop criteria
Parameters for the definition of the stop criteria.

Command variables
The following command variables are generated by the ANALOG IN
command in the method run and can be used in formulas under the des-
ignation 'Command name.Port name.Variable identification':

Identification

Description

BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

FIN

Command status:

1 = Command has ended at least once

0 = Command has not yet ended

not valid (variable not available) = Command
has not yet been started

UNI

Mapping unit which is defined for an IO port in the
configuration

[VAL]

Value of the command variables (optional, i.e.
'Command name.Portname' = 'Command
name.Portname.VAL'
). The value returned corre-
sponds to the status of the IO port after the com-
mand is exited; 1 = active, 0 = inactive.