5 cp - post-treatment, Post-treatment – Metrohm viva 1.1 (ProLab) User Manual

Page 660

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

648

■■■■■■■■

viva 1.1 (for process analysis)

Measuring interval

Measuring interval for data recording during plating or during the OCP
measurement (see Glossary, page 1055).

Input range

0.1 - 60.0 s (Increment: 0.1)

Default value

1.0 s

Number of measuring points

Shows the number of measuring points.

NOTE

The actual potential measurement takes place during the last 100 ms of
the measuring interval.

5.5.2.11.4.5

CP - Post-treatment

Tab: Method

CP Properties... Post-treatment

Command name

Name of the command.

Entry

25 characters

Cleaning

Cleaning potential

Potential that is applied to the electrodes during the cleaning time. This
potential can be used e.g. for the electrochemical cleaning of solid-state
electrode surfaces that have been contaminated with reaction products of
electrode redox processes.

Input range

–5.000 - 5.000 V (Increment: 0.001)

Selection

off

Default value

off

Cleaning time

Waiting time during which the cleaning potential is applied to the elec-
trodes. The field can only be edited for Cleaning potentialoff.

Input range

0.0 - 9,999.9 s (Increment: 0.1)

Default value

0.0 s