3 858 professional sample processor, 1 858 professional sample processor - overview, 858 professional sample processor – Metrohm viva 1.1 (ProLab) User Manual

Page 862: 858 professional sample processor - overview

Advertising
background image

7.3 858 Professional Sample Processor

■■■■■■■■■■■■■■■■■■■■■■

850

■■■■■■■■

viva 1.1 (for process analysis)

If it is found during monitoring that the validity period has expired, then
one of the following actions will be triggered automatically during the
start test:

Selection

Document message only | Display and docu-
ment message
| Do not start the determina-
tion

Default value

Display and document message

Document message only
The message that the validity period has expired will be automatically
saved with the determination.
Display and document message
A message is displayed and you can select whether you want to con-
tinue the run or cancel it. If the run is continued then the message that
the validity period has expired will be automatically saved with the
determination.
Do not start the determination
The determination is not started. The following message must be con-
firmed with [OK].

7.3

858 Professional Sample Processor

7.3.1

858 Professional Sample Processor - Overview

Dialog window: Configuration

Devices [Edit] Properties... Properties - 'De-

vice type' - 'Device name'

Model versions

NOTE

The 858 Professional Sample Processor is available in various ver-
sions that differ in the combination of integrated modules. The 858
Professional Sample Processor
is described in the online help with
all possible modules.

Different versions are available for different fields of application:

858.0010 Professional Sample Processor
Sample changer without peristaltic pump.

858.0020 Professional Sample Processor
Sample changer with integrated two-channel peristaltic pump.

858.0030 Professional Sample Processor
Sample changer with integrated two-channel peristaltic pump and
injection valve.

Advertising