2 coefficient calc 1, 3 coefficient calc 2 – Metrohm viva 1.1 (ProLab) User Manual

Page 722

Advertising
background image

5.6 Evaluation subwindow

■■■■■■■■■■■■■■■■■■■■■■

710

■■■■■■■■

viva 1.1 (for process analysis)

Remarks

The units of the coefficients are dependent on the units of the x and y
axes of the corresponding calibration curve.

5.6.7.4.2

Coefficient CALC 1

Variable
name

RS.’Command name’.’Substance name’.CALC1

Description

First-order coefficient of the calibration curve or standard addition curve.

Decimal pla-
ces

0…3…6

Unit

Calibration method DT:

L

⁻¹

Calibration method RC:

Concentration unit of the standard solu-
tion
= #g/L

L

· g⁻¹

Concentration unit of the standard solu-
tion
= #L/L

dimensionless

Concentration unit of the standard solu-
tion
= #mol/L

L

· mol⁻¹

Additional calibration methods:

Concentration unit of the standard solu-
tion
= #g/L

A

· L · g⁻¹ (for Evaluation quantity = Height)

C

· L · g⁻¹ (for Evaluation quantity = Area)

Concentration unit of the standard solu-
tion
= #L/L

A (for Evaluation quantity = Height)

C (for Evaluation quantity = Area)

Concentration unit of the standard solu-
tion
= #mol/L

A

· L · mol⁻¹ (for Evaluation quantity =

Height)

C

· L · mol⁻¹ (for Evaluation quantity = Area)

Remarks

The units of the coefficients are dependent on the units of the x and y axes of the corresponding
calibration curve.

5.6.7.4.3

Coefficient CALC 2

Variable
name

RS.'Command name'.'Substance name'.CALC2

Description

Second-order coefficient of the calibration curve or standard addition curve.

Decimal pla-
ces

0…3…6