2 evaluation - general, 1 general - general, 2 general - data processing – Metrohm viva 1.1 (ProLab) User Manual

Page 663: Evaluation - general

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1 (for process analysis)

■■■■■■■■

651

5.6.2

Evaluation - General

5.6.2.1

General - General

Subwindow section: Method

Evaluation General

Application
The parameters that are to be used for the processing of the acquired
data are defined in the General section. These parameters must be set
separately for each VA measuring command.

Selection of the measuring command

Measuring command

Selection list with all of the VA measuring commands with data acquisi-
tion defined in the method.

Selection

'VA measuring command'

Tabs

Data processing

Curve evaluation

5.6.2.2

General - Data processing

Tab: Method

Evaluation General Data processing

Smoothing

Measure for the curve smoothing used for the data processing. The higher
the smoothing that is selected, the more the measurement curves will be
smoothed.

CVS, CP
Input range

1 - 40

Default value

1

CPVS
Input range

1 - 40

Default value

10

Reversed peaks

on | off (Default value: off)

Reversed peaks are evaluated only if this check box is activated. Peaks that
exhibit a positive signal during measurement in cathode direction or a
negative signal during measurement in anode direction are referred to as
reversed peaks.