5 Method

viva 1.1 (for process analysis)

■■■■■■■■

461

Dosing unit

Selection of the dosing unit with which the sample solution is to be dosed
in.

Selection

'Name of the dosing unit' | not defined

Default value

not defined

'Name of the dosing unit'
Alphabetically organized selection of the dosing units from the table of
dosing units in the configuration.
not defined
The dosing unit must be assigned with not defined by the user at the
start of the method.

Dosing rate

Rate (volume per time unit) of ejection. The maximum rate depends on
the cylinder volume of the dosing unit used and on the diameter of the
tubing at the selected dosing port.

Input range

0.01 - 166.00 mL/min

Default value

2.00 mL/min

Selection

maximum

CAUTION

The maximum permitted dosing rate of 4 mL/min may not be exceeded
for capillaries with a diameter < 1 mm.

Filling rate

Rate (volume per time unit) of filling. The maximum rate depends on the
cylinder volume of the dosing unit used and on the diameter of the tubing
at the selected fill port.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

5.5.2.4.6

LQH

5.5.2.4.6.1

LQH - Overview

Dialog window: Method

LQH Properties... LQH - 'Command name'

Command for complex dosing tasks with a dosing device of the Dosino
800 type.