5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

610

■■■■■■■■

viva 1.1 (for process analysis)

Identification

Description

ELT

Variable for electrode test:

0 = Electrode test has not been performed

1 = Electrode test has been performed once

2 = Electrode test has been performed twice

3 = Electrode test has been performed three times

FIN

Command status:

1 = Command has been called and ended at
least once

0 = Command has been called and never
ended

invalid (variable not available) = Command has
never been called

HR

Highest current measuring range used:

1 = 224 mA | 3 = 20 mA | 5 = 2 mA | 7 = 200 µA |
9 = 20 µA | 11 = 2 µA | 13 = 200 nA | 15 = 20 nA |
17 = 2 nA | 19 = 200 pA

LR

Lowest current measuring range used:

19 = 200 pA | 17 = 2 nA | 15 = 20 nA | 13 = 200 nA
| 11 = 2 µA | 9 = 20 µA | 7 = 200 µA | 5 = 2 mA |
3 = 20 mA | 1 = 224 mA

NMP

Number of measuring points in measuring point list

RPM

Stirring rate in min

–1

STY

Type of stop with which the command was stopped:
1 = finished normally, 0 = stopped manually
with
[Quit] or [Stop]

Result variables if called by CALL VMS
The following result variables are generated in the method run when a
CVS command is called by CALL VMS and can be used in formulas:

Result variables per substance, VMS and replication:

Designation 'RS.'Command name'.'Substance
name'.VMS.REP{y}.Variable identification'

Identification

Description

SUBST

Substance name