5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

460

■■■■■■■■

viva 1.1 (for process analysis)

Addition volume #

Size of the individual volume of sample that is to be dosed in. The number
of addition volumes is defined by the amount defined in Number.

This parameter can also be modified live after the first time the command
has been executed.

Input range

0.010 - 1,000.000 mL (Increment: 0.001)

Default value

0.010 mL

Addition
Selection of the addition mode for the sample.

Selection

Add manually | Add with dosing device

Default value

Add manually

Add manually
Manual addition of the sample, at which time the track is stopped. A
message appears and an acoustic signal is emitted.
Add with dosing device
Automatic addition of the solution with a dosing device. The sample is
added through the Dosing port 1 defined in the configuration for the
dosing unit (see Chapter 6.6.3.2, page 785).

Message
With manual addition, the running of the track in which the command is
located is stopped and a message is displayed.

Only editable for Addition = Add manually, otherwise disabled (includ-
ing text input field).

Selection

Display standard message | Display message
defined by the user

Default value

Display standard message

Display standard message
If this check box is activated, then the standard message for additions
will be displayed.
Display message defined by the user
If this check box is activated, then the message defined in the text field
will be displayed.

Dosing device
Parameters for automatic dosing. This section is displayed only when
Addition = Add with dosing device.