5.6 Evaluation subwindow

■■■■■■■■■■■■■■■■■■■■■■

662

■■■■■■■■

viva 1.1 (for process analysis)

Manual
If this option is selected, then the base points can be specified man-
ually for the baseline calculation.

CPVS
Selection

Automatically | Manual

Default value

Manual

Automatically
If this option is selected, then the base points are determined automat-
ically for the baseline calculation.
Manual
If this option is selected, then the base points can be specified man-
ually for the baseline calculation.

Start base point

Manually specified start base point at which the baseline is to begin.

CVS
Input range

–5.000 - 5.000 V

Default value

0.0 V

CPVS
Input range

0 - t s

Default value

3.9 s
t is the total duration of data recording for curve
evaluation. It is the sum of plating time + stripping
time 1 + stripping time 2 + stripping time 3.

End base point

Manually specified end base point at which the baseline is to end.

CVS
Input range

–5.000 - 5.000 V

Default value

0.0 V

CPVS
Input range

0 - t s

Default value

6.5 s
t is the total duration of data recording for curve
evaluation. It is the sum of plating time + stripping
time 1 + stripping time 2 + stripping time 3.