4 Database

viva 1.1 (for process analysis)

■■■■■■■■

287

Recalculating determinations with calibration method DT or
RC

– Determinations of the sample type Sample can be recalculated

with a calibration. However, the determinations and the calibra-
tion must have been carried out with the same method. The
method version may differ.

– In order for the current calibration data to be used, viva auto-

matically recalculates the determination of the sample type
Standard first after the determinations have been selected.

– In order for determinations of the sample type Sample to be

reprocessed with a new, recalculated calibration, the selection of
determinations may contain only one determination of the sam-
ple type Standard.

– If a determination of the sample type Standard has been recal-

culated, then a prompt will appear asking whether the recalcu-
lated calibration data is to be saved and used for calculating the
substance concentration(s) in future determinations.
If you click on [Yes], the calibration data of the recalculated
standard will be used for calculating the substance concentra-
tion(s) in future determinations of the sample type Sample.
If you click on [No], the calibration data of the recalculated stan-
dard will not be made available for future determinations of the
sample type Sample.

– Determinations of the sample type Sample can be recalculated

without requiring a determination of the sample type Standard
to be selected. In this case, the original calibration data will be
used.