3 meas tmf - measuring parameters – Metrohm viva 1.1 (ProLab) User Manual

Page 554

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

542

■■■■■■■■

viva 1.1 (for process analysis)

Sensor

Temperature measurement

Type of temperature measurement

Selection

continuous | off

Default value

continuous

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Measuring parameters tab will be
used.

Stirrer

Stirrer

Selection of the stirrer.

Selection

1 | off

Default value

1

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate.

Input range

0 - 15

Default value

3

Switch off automatically

on | off (Default value: on)

If this check box is activated, the stirrer will be switched off automatically
when the command has finished.

5.5.2.7.6.3

MEAS TMF - Measuring parameters

Tab: Method

MEAS TMF Properties... Measuring parameters

Command name

Name of the command.

Entry

25 characters