5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
498
■■■■■■■■
viva 1.1 (for process analysis)
Signal pattern
Selection of a signal pattern predefined in the Configuration program
part.
The input lines (inputs) are always numbered from right to left:
Input 7 6 5 4 3 2 1 0
Bit
7 6 5 4 3 2 1 0
Example:
*******1 expects an active input line 0 (1 = set).
NOTE
Input lines that are of no interest or for which no defined condition can
be predicted should be masked with an asterisk *.
Timeout
on | off (Default value: off)
If this check box is activated, then a maximum waiting time can be
entered for the input signal.
Input range
0.0 - 9,999.9
Default value
10.0
Selection
min | s
Default value
min
NOTE
If the Timeout check box is not activated, then there will be an infin-
itely long wait for the input signal.
5.5.2.6.4
SEND
5.5.2.6.4.1
SEND - Overview
Dialog window: Method
▶ SEND ▶ Properties... ▶ SEND - 'Command name'
Command for sending event messages to commands, tracks or to
viva.
Appearance
The command has the following appearance: