4 dosing - filling, 5 dosing - emptying, 6 dosing - dosing a fixed volume – Metrohm viva 1.1 (ProLab) User Manual

Page 973: Filling, Emptying, Fixed volume

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

8 Manual control

viva 1.1 (for process analysis)

■■■■■■■■

961

[Stop]

Stops preparation for the selected dosing device(s).

8.4.2.4

Dosing - Filling

Program part: Manual control

The filling of dosing units can be started and stopped here. This tab
appears both when a single dosing device or all dosing devices are
selected.

[Start]

Starts filling for the selected dosing device(s).

[Stop]

Stops filling for the selected dosing device(s).

8.4.2.5

Dosing - Emptying

Program part: Manual control

Emptying of dosing units can be started and stopped here. This tab
appears both when a single dosing device or all dosing devices with dos-
ing units are selected.

[Start]

Starts emptying for the selected dosing device(s). In the process, the
parameters defined for the dosing units will be used.

[Stop]

Stops emptying for the selected dosing device(s).

8.4.2.6

Dosing - Dosing a fixed volume

Program part: Manual control

The dosing of a preset volume can be started and stopped here. This tab
only appears if a single dosing device is selected.

Volume

Fixed volume that is to be dosed.

Input range

0.00 - 99,999.9 mL

Default value

1.000 mL

Dosing rate

The volume is dosed at this rate. The maximum dosing rate depends on
the cylinder volume of the dosing unit used. If the entered dosing rate is
too high for the selected dosing device, it will automatically be reduced
during dosing to the largest possible value.