5 Method

viva 1.1 (for process analysis)

■■■■■■■■

645

Reference electrode

Selection of an electrode of the Reference electrode type from the elec-
trodes available in the electrode table.

Selection

Reference electrode | 'Electrode name'

Default value

Reference electrode

Reference electrode
Standard electrode of the Reference electrode type, which cannot
be deleted.
'Electrode name'
Electrode of the Reference electrode type from the electrode table.

Auxiliary electrode

Selection of an electrode of the Auxiliary electrode type from the elec-
trodes available in the electrode table.

Selection

Auxiliary electrode | 'Electrode name'

Default value

Auxiliary electrode

Auxiliary electrode
Standard electrode of the Auxiliary electrode type, which cannot be
deleted.
'Electrode name'
Electrode of the Auxiliary electrode type from the electrode table.

Electrode check

The electrode test for the reference and auxiliary electrodes can be acti-
vated and deactivated only in the START command (see "Electrode test
for VA measuring commands", page 579)
.

Stirrer

Stirring rate

Number of revolutions per minute of the stirrer.

Input range

0 - 3,000 min

⁻¹ (Increment: 100)

Default value

2,500 min

⁻¹

Hydrodynamic measurement

on | off (Default value: on)

If this check box is activated, then stirring will continue during the sweep.

Only editable for Sensor type = RDE, otherwise disabled.