5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

640

■■■■■■■■

viva 1.1 (for process analysis)

Principle
In the CP measuring mode with the option Measure galvanostatically
(see Glossary, page 1046), a constant current is applied to the working
electrode (see A) and the diffusion-controlled change in the electrode
potential over time is measured (see B).

In the CP measuring mode with the option Measure open-circuit
potential
(see Glossary, page 1055), the circuit is opened and the
potential between the working electrode and the reference electrode is
measured during the indicated measurement duration (see B).

t

tot

I

const

t

interval

t

meas

A

t

interval

t

meas

Measure galvanostatically:

t

meas

≘ Plating time

Measure open-circuit poten-
tial
:

t

meas

Measurement duration

(see Chapter 5.5.2.11.4.4, page
647)

t

interval

Measuring interval (see page
648)
.

U

t

B

Sequence of individual steps
In the CP command, the following individual steps are successively exe-
cuted in the method run for the device defined in the command: