5 series start, Series start – Metrohm viva 1.1 (ProLab) User Manual

Page 599

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1 (for process analysis)

■■■■■■■■

587

Live display

on | off (Default value: on)

If this check box is activated, then the track will be displayed in the live
display.

Return immediately

on | off (Default value: off)

If this check box is activated, then the program does not wait for this track
to be completed, but rather this track responds immediately to the track
with the call command. From now on, both tracks will be running simulta-
neously.

Command variables
The following command variables are generated by the command VA
TRACK
in the method run and can be used in formulas under the desig-
nation 'Command name.Variable identification':

Identifi-
cation

Description

BSY

Command status:

1 = BUSY, HOLD or ERROR

0 = READY

invalid (variable not available) = Command has never
been started

FIN

Command status:

1 = Command has ended at least once

0 = Command has not yet ended

invalid (variable not available) = Command has never
been started

5.5.2.9.5

SERIES START

Dialog window: Method

SERIES START Properties... SERIES START - 'Com-

mand name'

Start command for the series start track.

Appearance
The command has the following appearance:

Advertising