5 Method

viva 1.1 (for process analysis)

■■■■■■■■

713

Variable

Description

2

R

Coefficient of determination.

i

y

y value of calibration point i.

y

Mean value of all y values of the calibra-
tion points.

i

yˆ

The y value of calibration point i calcu-
lated with the calibration curve.

5.6.7.5

Volume results

5.6.7.5.1

Total volume

Variable name

RS.VTOT

Description

Current total volume in the measuring vessel.

Decimal places

0…3…6

Unit

#L

Remarks

The value of these variables is automatically set to 0 at the start of a
determination. The respective defined addition volume for the RS.VTOT
variable is added with each command ADD AUX (only if the option
Include volume in calculation is activated), ADD STD, ADD SAMPLE
and ADD SAMPLE DT.

With the SET TOTAL VOLUME command, any value can be manually
entered for the total volume in the measuring vessel (e.g. after automatic
rinsing or manual solution exchange).

5.6.7.5.2

Total volume per variation

Variable name

RS.’Command name’.VAR{x}.VTOT

Description

Current total volume in the measuring vessel for variation{x}.

Decimal places

0…3…6

Unit

#L

5.6.7.6

Concentration results

5.6.7.6.1

Substance concentration in the sample

Variable name

RS.’Command name’.’Substance name’.CONC

Description

Concentration of the substance in the sample.

Decimal places

0…2…6