2 analog out - general – Metrohm viva 1.1 (ProLab) User Manual

Page 573

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1 (for process analysis)

■■■■■■■■

561

Analog outputs

Command variables
The following command variables are generated by the ANALOG OUT
command in the method run and can be used in formulas under the des-
ignation 'Command name.Port name.Variable identification':

Identification

Description

BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

DEF

Default value which is defined in the configuration
for an IO port.

FIN

Command status:

1 = Command has ended at least once

0 = Command has not yet ended

not valid (variable not available) = Command
has not yet been started

UNI

Mapping unit which is defined for an IO port in the
configuration

[VAL]

Value of the command variables (optional, i.e. 'Com-
mand name.Portname'
= 'Command
name.Portname.VAL'
). The value returned corre-
sponds to the status of the IO port after the com-
mand is exited.

5.5.2.8.4.2

ANALOG OUT - General

Tab: Method

ANALOG OUT Properties... ANALOG OUT - 'Command

name'

General

Command name

Name of the command.

Entry

25 characters

Device

Device name

Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.

Selection

'Device name' | not defined

Default value

not defined