8 port, 1 port - overview, Port – Metrohm viva 1.1 (ProLab) User Manual

Page 586

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

574

■■■■■■■■

viva 1.1 (for process analysis)

5.5.2.8.8

PORT

5.5.2.8.8.1

PORT - Overview

Dialog window: Method

PORT Properties... PORT - 'Command name'

Command for changing the port position on a multi-port valve.

Instruments
This command can be executed with the following instruments:

Titrando: 901, 902, 904, 905, 906, 907

Dosing Interface: 846

Appearance
The command has the following appearance:

Parameters
The parameters for the PORT command are configured in the following
dialog window:

PORT - Properties

Command variables
The following command variables are generated by the PORT command
in the method run and can be used in formulas under the designation
'Command name.Variable identification':

Identification

Description

BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

EPP

Port position after exiting the command: 0 = invalid
position

FIN

Command status:

1 = Command has ended at least once

0 = Command has not yet ended

not valid (variable not available) = Command
has not yet been started

Advertising