Metrohm viva 1.1 (ProLab) User Manual
Page 549
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
537
089 Photometer
Input range
0.25 - 999,999.00 s (Increment: 0.25)
Default value
2.0 s
Stop measured value
Stop when the entered measured value has been reached since the start
of the measurement.
Input range
0 - 1E+10 (max. 10 digits)
Default value
0.0
Selection
off
Default value
off
off
No stop.
Measurement without drift control
If the Measurement without drift control option has been selected,
then the following parameters will be displayed:
Max. waiting time
Maximum period of time for measurement.
Input range
0 - 999,999 s
Default value
120 s
Measuring interval
Time interval for entering a measured value in the measuring point list.
Avantes Spectrometer
Input range
0.1 - 999,999.0 s (Increment: 0.1)
Default value
2.0 s
089 Photometer
Input range
0.25 - 999,999.00 s (Increment: 0.25)
Default value
2.0 s
Stop measured value
Stop when the entered measured value has been reached since the start
of the measurement.
Input range
0 - 1E+10 (max. 10 digits)
Default value
0.0
Selection
off
Default value
off
off
No stop.