5 Method

viva 1.1 (for process analysis)

■■■■■■■■

625

Sweep

The following steps are repeated with the specified number of preparation
cycles and measuring cycles.

1 Equilibration potential

Equilibration potential is applied to the electrodes.

2 Equilibration time

Idle until the period defined in the Equilibration time parameter
has expired.

3 Plating potential

Plating potential is applied to the electrodes.

4 Plating time

Idle until the period defined in the Plating time parameter has
expired.

5 Stripping potential 1

Stripping potential 1 is applied to the electrodes.

6 Stripping time 1

Idle until the period defined in the Stripping time 1 parameter has
expired.

7 Stripping potential 2

Stripping potential 2 is applied to the electrodes.

8 Stripping time 2

Idle until the period defined in the Stripping time 2 parameter has
expired.

9 Stripping potential 3

Stripping potential 3 is applied to the electrodes.

10 Stripping time 3

Idle until the period defined in the Stripping time 3 parameter has
expired.