Metrohm viva 1.1 (ProLab) User Manual
Page 637
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
625
Sweep
The following steps are repeated with the specified number of preparation
cycles and measuring cycles.
1 Equilibration potential
Equilibration potential is applied to the electrodes.
2 Equilibration time
Idle until the period defined in the Equilibration time parameter
has expired.
3 Plating potential
Plating potential is applied to the electrodes.
4 Plating time
Idle until the period defined in the Plating time parameter has
expired.
5 Stripping potential 1
Stripping potential 1 is applied to the electrodes.
6 Stripping time 1
Idle until the period defined in the Stripping time 1 parameter has
expired.
7 Stripping potential 2
Stripping potential 2 is applied to the electrodes.
8 Stripping time 2
Idle until the period defined in the Stripping time 2 parameter has
expired.
9 Stripping potential 3
Stripping potential 3 is applied to the electrodes.
10 Stripping time 3
Idle until the period defined in the Stripping time 3 parameter has
expired.