2 control - properties – Metrohm viva 1.1 (ProLab) User Manual

Page 579

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1 (for process analysis)

■■■■■■■■

567

Parameters
The parameters for the CONTROL command are configured in the fol-
lowing dialog window:

CONTROL - Properties

Command variables
The following command variables are generated by the CONTROL com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':

Identification

Description

BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

FIN

Command status:

1 = Command has ended at least once

0 = Command has not yet ended

not valid (variable not available) = Command
has not yet been started

IVS

Value of the input at the start of the command.

IVE

Value of the input at the end of the command.

OVS

Value of the output at the start of the command.

Direct controller/2-point controller command status:
1 = active, 0 = inactive

PID/PWM: -100 to 100% (value of the analog output
or the pulse frequency)

OVE

Value of the output at the end of the command.

Direct controller/2-point controller: 1 = active, 0 =
inactive

PID/PWM: -100 to 100% (value of the analog output
or the pulse frequency)

5.5.2.8.6.2

CONTROL - Properties

Dialog window: Method

CONTROL Properties... CONTROL - 'Command

name'

Command name

Name of the command.

Entry

25 characters