5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
626
■■■■■■■■
viva 1.1 (for process analysis)
11 Stripping potential 4
Stripping potential 4 is applied to the electrodes.
12 Stripping time 4
Idle until the period defined in the Stripping time 4 parameter has
expired.
13 Stripping potential 5
Stripping potential 5 is applied to the electrodes.
14 Stripping time 5
Idle until the period defined in the Stripping time 5 parameter has
expired.
15 Cleaning potential
Cleaning potential is applied to the electrodes.
16 Cleaning time
Idle until the period defined in the Cleaning time parameter has
expired.
Post-treatment
only for hydrody-
namic measurement
= on
1 Stirrer off
The stirrer is switched off.
only for standby
potential = off
2 Potential off
The potential applied to the electrodes is switched off.
only for standby
potential
≠
off
3 Standby potential on
Standby potential is applied to the electrodes.
Parameters
The parameters for the CPVS command are set on the following tabs:
■
■
■
■
■